Power

115 V / 60 Hz, single phase[2]

Vacuum

P2 pump[2]

How it works

The Thermoflex 900 uses an air-cooled condenser, a refrigeration system, and a recirculating pump to maintain fluid temperature and volume. The Thermoflex 900 includes a microprocessor controller for operating control.[2][3]

The Thermoflex 900 can include a positive-displacement P1 pump or a positive-displacement P2 pump. The P1 pump is stated at 2.1 gpm at 60 psig, and the P2 pump is stated at 4.0 gpm at 60 psig.[3][2]

Where it fits in the process flow

The Thermoflex 900 supplies temperature-controlled fluid to external circulation applications, open containers, and closed-loop systems.[2][3]

The Thermoflex 900 uses an air-cooled condenser that removes application heat with ambient room air.[2][3]

What do the numbers mean?

Power & electrical1

Power requirement
115 V / 60 Hz, single phase[2]
Accurate?

Vacuum & pumping4

Pump configuration
P2 pump[2]
Accurate?
Pump performance
4.0 gpm at 60 psig[2]
Accurate?
Pump configuration
P1 pump[3]
Accurate?
Pump performance
2.1 gpm at 60 psig[3]
Accurate?

Configuration & options11

Manufacturer
Thermo Scientific[1]
Accurate?
Model
Thermoflex 900[1]
Accurate?
Type
Recirculating Chiller[1]
Accurate?
Cooling method
Air cooled only[2]
Accurate?
Cooling capacity
900 W (3074 BTU) at 20°C[2]
Accurate?
Reservoir volume
7.2 liter (1.9 gallon)[2]
Accurate?
Set point temperature range
+5°C to +40°C[2]
Accurate?
Ambient temperature range
+10°C to +40°C[2]
Accurate?
Refrigerant
R134A, 19.0 oz[2]
Accurate?
Communications
RS232 ports on the back[3]
Accurate?
Compliances
SEMI S2 Compliant – S2-0703, S8-0705, S14-0704 & F47-0706[2]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
Thermo Scientific Neslab ThermoFlex 900 Recirculating Chiller, AIR COOLED, P2 Pump, 900 Watt, 115 VAC——Air cooled, P2 pump, 900 Watt, 115 VAC.idealvac.com[2]
Thermo Scientific Neslab ThermoFlex 900 Recirculating Chiller, AIR COOLED, P1 Pump, 900 Watt, 115 VAC——Air cooled, P1 pump, 900 Watt, 115 VAC.idealvac.com[3]
ThermoFlex 900 Recirculating Chiller, AIR COOLED, P2 Pump——P2 pump; 4.0 gpm at 60 psig.idealvac.com[2]
ThermoFlex 900 Recirculating Chiller, AIR COOLED, P1 Pump——P1 pump; 2.1 gpm at 60 psig; RS232 ports on the back.idealvac.com[3]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Pump configurationP2 pump[2]
  • Pump performance4.0 gpm at 60 psig[2]
  • Pump configurationP1 pump[3]
  • Pump performance2.1 gpm at 60 psig[3]
  • Power requirement115 V / 60 Hz, single phase[2]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What type of cooling is used?

The Thermoflex 900 is air cooled, and the condenser uses ambient room air to remove application heat.[2][3]

Not publicly documented

The following facts about the Thermoflex 900 Recirculating Chiller are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the Thermoflex 900 Recirculating Chiller are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the Thermoflex 900 Recirculating Chiller are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the Thermoflex 900 Recirculating Chiller are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the Thermoflex 900 Recirculating Chiller is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the Thermoflex 900 Recirculating Chiller are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (3)Every fact above is drawn from these public sources
  1. [1]Equipment inventory listing
  2. [2]idealvac.com — idealvac.comidealvac.com
  3. [3]idealvac.com — idealvac.comidealvac.com
Was this page accurate?

Last updated Oct 6, 2026.

Compare with similar tools

View all →