Electroglas 1034, Keysight 4142 B, side by side. Every value shown is drawn from the cited encyclopedia entries.

Side-by-side comparison of selected equipment models
Attribute
1034Electroglas
4142 BKeysight
OEMElectroglasKeysight
CategoryTest & ProbeTest & Probe
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
FamilyElectroglas 1034Keysight 4142 B
Specifications
Architecture—Mainframe with 8 slots for plug-in measurement units[1]
Integrated measurement functions—Current/voltage sources, nanovoltmeters, and picoammeters in one instrument[1]
Source Monitor Unit (SMU) sensitivity—Forces voltage while measuring currents down to 20 fA; forces current while monitoring voltages down to 40 µV[1]
HP 41420A High Power SMU—2 slots; 40 µV to 200 V, 20 fA to 1 A (±200 V and ±1 A)[1]
HP 41421B Medium Power SMU—1 slot; 40 µV to 100 V, 20 fA to 100 mA (±100 V and ±100 mA)[1]
HP 41422A High Current Unit (HCU)—2 slots; pulse mode only; 100 µs pulsed measurements up to ±10 V and 10 A[1]
HP 41423A High Voltage Unit (HVU)—2 slots; DC or pulsed measurements to ±1000 V; low-leakage design measures currents down to 2 pA[1]
HP 41425A Analog Feedback Unit (AFU)—1 slot; used with two SMUs for analog voltage/current searches; finds target values in 12 ms[1]
Maximum configuration coverage—2000 V max using two HVUs; 20 A (pulsed) using two HCUs[1]
Maximum output power—Sum of all measurement units' instantaneous output power cannot exceed 32 W[1]
Maximum common-to-ground voltage—±42 V[1]
Ground Unit (GNDU)—High-current SMU set to 0 V, used as measurement ground; Kelvin connection; maximum offset voltage ±500 µV; current range ±1.6 A[1]
Program memory—Approx. 2000 program steps, groupable into 99 subroutines[1]
Data memory—Maximum 4095 data values (binary) or 1023 data values (ASCII)[1]
Measurement modes—Plug-in units operate individually or in synchronized modes to perform ten different types of measurements[1]
Interfaces—HP-IB; external trigger input (TTL, negative logic, minimum pulse width 100 µs); external trigger output (TTL, negative logic)[1]
Self-test and calibration—Self-test at power-up; offset errors in each measurement unit auto-calibrated every 30 minutes[1]
Operating temperature—5 °C to 40 °C[1]
Operating humidity—5% to 80% RH[1]
Storage temperature and humidity—-40 °C to 65 °C; <90% RH at 65 °C[1]
Power requirements—100/120/220 V ±10%; 240 V -10%/+5%; maximum 750 VA (48-66 Hz)[1]
Dimensions—426 mm W × 235 mm H × 676 mm D[1]
Weight—Mainframe approx. 23 kg; plug-in units 2-3 kg each[1]
Typical measurement time (SMU)—HP 41420A/HP 41421B SMU (20 V/100 mA range, spot measurement): force I or V in 3.5 ms; measure I or V in 4.0 ms[1]